Transferpettor Positive Displacement Pipette FAQ


Browse through some common questions about the Transferpettor Positive Displacement Pipette:

1.
My Transferpettor™ always dispenses considerably smaller volume than it’s setting. What’s wrong?
2.
Are the tips and seals for the Transferpettor™ autoclavable?



1.
My Transferpettor™ always dispenses considerably smaller volume than it’s setting. What’s wrong?
Before you send it out for calibration, you may want to make sure you the tips are placed on the pipette correctly. The pipetting button must also be pushed down before locking the tip into place.

If the button is not pushed down all the way, the tip can be pushed on further that appropriate, and the volume will be smaller. If the tips are put on correctly, and the volume is still inaccurate, the Transferpettor™ can have a calibration adjustment performed. However, this is not a user function. The pipette should be sent in for authorized service.

Return to Top

2.
Are the tips and seals for the Transferpettor™ autoclavable?
No.

Return to Top